Mass spectrometry & spectroscopy
Monitoring disinfection byproducts in drinking water: A sensitive IC-MS/MS approach
Aug 05 2025
While chlorination is essential for controlling pathogens in drinking water, it can produce harmful byproducts, including haloacetic acids (HAAs), dalapon, and bromate, some of which are potentially carcinogenic. Regulatory bodies like the US EPA and the EU have set a maximum contamination limit of 60 µg/L for the sum of five key HAAs (HAA5: MCAA, MBAA, DCAA, DBAA, TCAA).
EPA Method 557 outlines a procedure for quantifying these compounds at trace levels across diverse water types. In this method, a Metrohm ion chromatograph (IC) is paired with an Agilent triple quadrupole mass spectrometer (MS), enabling direct, high-sensitivity analysis without the need for sample extraction. The Metrohm Suppressor Module minimises eluent interference, while the Metrosep A Supp 19 column ensures excellent separation of analytes from matrix components.
Even in complex water samples, matrix spike recoveries at 1 µg/L ranged from 65–115%, and minimum reporting levels (MRLs) fell between 0.025 and 0.25 µg/L.This robust IC-MS/MS method fully complies with EPA Method 557 and provides a reliable tool for safeguarding drinking water quality.
More information online
Digital Edition
Lab Asia Dec 2025
December 2025
Chromatography Articles- Cutting-edge sample preparation tools help laboratories to stay ahead of the curveMass Spectrometry & Spectroscopy Articles- Unlocking the complexity of metabolomics: Pushi...
View all digital editions
Events
Jan 21 2026 Tokyo, Japan
Jan 28 2026 Tokyo, Japan
Jan 29 2026 New Delhi, India
Feb 07 2026 Boston, MA, USA
Asia Pharma Expo/Asia Lab Expo
Feb 12 2026 Dhaka, Bangladesh