Laser diffraction technology: An innovative cell for "undiluted" measurements of concentrated samples
Apr 03 2019
Laser diffraction technologies are known as accurate and fast technologies to characterise the size distribution of sample from few tens of nm to several thousand of µm, for emulsions, suspensions or powders. In this technology, particles should scatter individually light and distance between particles should be sufficient to prevent multiscattering regimes.
For wet dispersions, due to this method principle, we generally need to dilute samples strongly as soon as concentration is high.
HORIBA launches a new cell for the LA960 particle size analyser dedicated to any analysis of «concentrated samples with low viscosity".
This cell, composed of two windows, positioned on a cell holder integrates a spacer, with variable thickness (selectable by user), to adjust the distance between these two windows.
Samples concentrated with low viscosity are introduced in this new cell type by using a syringe. When filled up, cell is placed inside its holder and recognised automatically by the laser diffraction device to start runs.
This unique concept, of windows and spacer with variable thicknesses, allows to adjust the optical path travelled by light source and thus, controls, reduces or avoids dilutions requested from samples highly concentrated.
For applications such as inks, emulsions or pigment suspensions, it is requested by users to avoid dilutions as these dilutions may modify/change the sample formulation and / or underestimate the presence of agglomerates.
Here below an example of results for suspensions used for battery electrodes manufacture.
With a standard circulation cell, dilution is compulsory, and the aggregates are not detected, as statistically not taken during sampling.
With the use of the high concentration cell, no dilution is applied, and the aggregates are then analysed properly. This new cell allows to work 4000 times more concentrated than with classical circulation wet cells. More representative samples can de definitively analysed in such conditions and avoid sampling errors.
This example shows the interest of using such cell for better detection capabilities and full sample integrity.
Digital Edition
International Labmate Buyers' Guide 2024/25
June 2024
Buyers' Guide featuring: Product Listings & Manufacturers Directory Chromatography Articles - Enhancing HPLC Field Service with fast-response, non-invasive flowmeters - Digital transformatio...
View all digital editions
Events
Jul 07 2024 Dublin, Ireland
Jul 20 2024 Denver, CO, USA
Jul 21 2024 Cape Town, South Africa
Jul 28 2024 San Diego, CA USA
Jul 30 2024 Jakarta, Indonesia