• Automated, One-Step SPE and Concentration of Nitrogen and Phosphorus Containing Pesticides in Water
    PowerPrep SPE and PowerVap Concentrator systems

    Laboratory products

    Automated, One-Step SPE and Concentration of Nitrogen and Phosphorus Containing Pesticides in Water

    Introduction
    Nitrogen and phosphorus containing pesticides are being widely used as replacements for organochlorine pesticides. Initially considered more desirable than organochlorine pesticides because of the way they rapidly break down in the environment, they inhibit the enzyme acetylcholine-sterase and disrupt nerve function. Nitrogen and phosphorus containing pesticides, particularly organo-phosphates have proven to be extremely toxic to vertebrates.

    Despite their ability to breakdown rapidly, nitrogen and phosphorus containing pesticides constitute a major health concern due to their extreme toxicity and their tendency to persist in drinking water. The following procedure outlines the use of Solid Phase Extraction (SPE) to extract phosphorus and nitrogen containing pesticides for analysis by GC via EPA Methods 507, 614, 1657 and 8141.

    Instrumentation & Consumables
    FMS, Inc. PowerPrep™ SPE System
    FMS, Inc. PowerVap™ Concentrator
    FMS, Direct-to-Vial concentrator tubes
    Waters, 1 gm Oasis HLB® Cartridge
    Varian Chem Elut Hydromatrix® cartridge
    Thermo Trace GC w/Polaris QMS
    PowerPrep SPE
    The HLB Cartridge is conditioned with 10 mL methanol
    The HLB Cartridge is conditioned with 10 mL DI H2O
    Sample is loaded across HLB Cartridge via vacuum
    The sample bottle is auto-rinsed and rinse loaded on HLB
    The cartridge is dried with nitrogen
    The cartridge is eluted with ethyl acetate followed by methylene chloride, passed across a Chem Elut Hydromatrix® cartridge and eluted directly into the Direct-to-Vial FMS PowerVap.

    PowerVap Concentrator
    Pre-heat temp: 55 °C
    Pre-heat time: 30 minutes
    Heat in Sensor mode: 55 °C
    Nitrogen Pressure: 15 PSI

    Procedure: Sample Prep and Extraction

    1. Two, 1 liter water samples spiked using Ultra NPM-525C & NPM 525B Nitrogen/Phosphorus Pesticide Mix (Diluted).
    2. The samples are acidified to PH <2.
    3. Samples are equilibrated for 15 minutes.
    4. Two samples are loaded onto to corresponding sample positions on the FMS PowerPrep SPE system.
    5. The Program is initiated to run each sample sequentially.
    6. The sample is extracted and automatically transferred to the FMS PowerVap Concentrator system with Direct-to-Vial tubes.
    7. Extracts are concentrated with the PowerVap system to <1 mL.
    8. Extracts are removed from the PowerVap system, volume adjusted to 1 mL, internal standard added (Ultra ISM-510) and transferred to Agilent GC for analysis.

    Results

      Position Position Mean  
    Compound #1 #2 Rec       RSD
    Alachlor 4.72 4.61 93.30% 2.40%
    Ametryn 4.46 4.5 89.60% 0.90%
    Atraton 4.59 4.26 88.50% 7.50%
    Atrazine 4.49 4.42 89.10% 1.60%
    Bromacil 4.16 5.07 92.50% 19.20%
    Butachlor 4.08 4.09 81.70% 0.20%
    Butylate 3.62 3.54 71.60% 2.20%
    Carboxin 4.3 3.64 79.40% 16.60%
    Chlorophram 4.61 4.17 87.80% 10.00%
    Cycloate 4.11 3.9 80.10% 5.20%
    Dichlorvos 4.23 3.91 81.40% 7.90%
    Diphenamid 4.13 4.55 86.80% -9.70%
    Disulfoton* 4.03 3.32 73.50% 19.30%
    EPTC 3.74 3.41 71.50% 9.20%
    Ethoprop 4.64 4.44 90.80% 4.40%
    Fenamiphos 4.21 5.07 92.80% 18.50%
    Fenarimol 5.27 4.82 100.90% 8.90%
    Fluridone 5.51 5.06 105.70% 8.50%
    Hexazinone   4.86 4.69 95.50% 3.60%
    Methyl Paraoxon 5.35 5.56 109.10% -3.80%
    Metolachlor 4.5 4.41 89.10% 2.00%
    Mevinphos 4.57 4.38 89.50% 4.20%
    MGK 264 4.13 3.59 77.20% 14.00%
    Molinate 4.24 3.97 82.10% 6.60%
    Napropamide 4.49 4.28 87.70% 4.80%
    Norflurazon 4.74 4.67 94.10% 1.50%
    Pebulate 4.07 3.52 75.90% 14.50%
    Prometon 4.16 4.3 84.60% -3.30%
    Prometryn 4.4 4.57 89.70% -3.80%
    Pronamide** 4.46 4.36 88.20% 2.30%
    Propazine 4.72 4.55 92.70% 3.70%
    Simazine 4.2 4.24 84.40% -0.90%
    Simetryn 4.79 4.73 95.20% 1.30%
    Stirofos 4.33 4.4 87.30% -1.60%
    Tebuthiuron 4.36 4.18 85.40% 4.20%
    Terbacil 4.81 5.33 101.40% 10.30%
    Terbufos** 3.58 3.51 70.90% 2.00%
    Terbutryn 4.43 4.82 92.50% -8.40%
    Triademefon 5.69 4.82 105.10% 16.60%
    Tricyclazole 5.25 4.95 102.00% 5.90%
    Vemolate 3.69 3.529 72.20% 4.50%


    Conclusions
    Analysis of duplicate samples across the automated, one-step PowerPrep SPE Direct-to-Vial Concentration system demonstrates highly efficient extractions with excellent recoveries and repro-ducibility. Watch the product video.

    For more information contact FMS:
    FMS Inc.
    580 Pleasant Street
    Watertown, MA 02472
    Phone: (617) 393-2396
    Fax: (617) 393-0194
    Email: onlineinfo@fms-inc.com
    Web site: fmsenvironmental.com
     


    Digital Edition

    Lab Asia 32.2 April

    April 2025

    Chromatography Articles - Effects of small deviations in flow rate on GPC/SEC results Mass Spectrometry & Spectroscopy Articles - Waiting for the present to catch up to the future: A bette...

    View all digital editions

    Events

    Medtec Japan 2025

    Apr 09 2025 Tokyo, Japan

    FORUMESURE

    Apr 22 2025 Hammamet, Tunisia

    Korea Lab 2025

    Apr 22 2025 Kintex, South Korea

    Analytica Anacon India & IndiaLabExpo

    Apr 23 2025 Mumbai, India

    Analitika Expo 2024

    Apr 23 2025 Moscow, Russia

    View all events