NMR Basics 101 — How to Tune a Probe

The probe is at the heart of an NMR spectrometer, in the centre of the applied magnetic field. The probe holds the:

  • Sample and sample spinning apparatus
  • Radio frequency (RF) coils
  • Temperature controller
  • Detector coils

The sample tube containing the sample is placed inside the probe. The sample tube can be various sizes depending on the type and size of the sample. An article showing the use of different probes for different samples is: Features Added To Mqc Benchtop NMR Analyser.

There are methods to shield the main magnetic field from the effects of the fields generated inside the probe. This can simply be an electric circuit.

Spinner and Temperature

The temperature circuit in the probe maintains a constant temperature; this is needed when the effects of temperature on the sample are being measured. The circuit monitors the sample temperature and adjusts the flow of heating or cooling agent, usually a gas, around the sample.

The purpose of the sample spinner is to rotate the NMR sample tube about its axis. In doing so, each spin in the sample located at a given position along the Z-axis and radius from the Z-axis, will experience the average magnetic field in the circle defined by this Z and radius. This procedure produces a narrow line width in the resulting spectrum.

Radio Frequency and Detector Coils

In many NMR spectrometers, the RF and detector coils are the same coils. These coils deliver a pulse of energy at the frequency needed for the nuclei under investigation. The same coils will also act as detectors for the RF signal from the sample.

For a defined NMR experiment, the probe needs to be tuned for the nucleus under consideration. Different nuclei need different frequencies, and so the probe will need retuning for each nucleus investigated. If a different sample matrix is used, a retune will be necessary, especially if a different solvent type is used to prepare the sample.

Signal-to-Noise Ratio

The signal-to-noise (S/N) ratio is a measure of an instrument’s sensitivity. S/N is proportional to the square root of the number of scans (or experiments). If the S/N ratio is reduced by a factor of 10 due to increased noise in the circuits, 100 more scans will have to be made to achieve the original sensitivity.

Tuning is important so that the spectrometer is operating at the most sensitive possible. The signal received from a sample is very small and so the noise in the circuits needs to be reduced to a minimum value.

Tuning a Probe

The term tuning involves two connected activities:

  • Tuning the coil to the required frequency
  • Matching the coil to the correct impedance ensuring its efficient operation

The circuit requiring tuning is composed of inductive (L) and capacitive (C) elements — the L and C values determine the frequency of the RF circuit. A circuit is tuned by adjusting the capacitor values until the required frequency is reached.

Image Source


Digital Edition

Lab Asia Dec 2025

December 2025

Chromatography Articles- Cutting-edge sample preparation tools help laboratories to stay ahead of the curveMass Spectrometry & Spectroscopy Articles- Unlocking the complexity of metabolomics: Pushi...

View all digital editions

Events

Smart Factory Expo 2026

Jan 21 2026 Tokyo, Japan

Nano Tech 2026

Jan 28 2026 Tokyo, Japan

Medical Fair India 2026

Jan 29 2026 New Delhi, India

SLAS 2026

Feb 07 2026 Boston, MA, USA

Asia Pharma Expo/Asia Lab Expo

Feb 12 2026 Dhaka, Bangladesh

View all events