Identifying Cannabinoids in Baked Goods with Minimal Sample Preparation

Thermo Fisher Scientific, Inc announced a new method for identifying marijuana cannabinoids in baked goods that is faster, easier-to-use and requires less sample preparation. Using ultra high performance liquid chromatography with mass spectrometry detection (UHPLC/MS), the application tests baked goods samples obtained from a local state police forensic laboratory and is detailed in the application note ‘Identification of Cannabinoids in Baked Goods by UHPLC/MS’.

The evidence is first screened for marijuana using simple chemical tests, such as the Duqenois-Levine test and then gas chromatography-mass spectrometry (GC/MS) is used to positively identify cannabinoids. This conventional technique works well for leaf marijuana, hashish and residue collected from smoking paraphernalia. UHPLC/MS however is an alternate method for identifying marijuana cannabinoids in complex food matrices such as baked goods.

Incorporating the Thermo Scientific Accela UHPLC system, the Thermo Scientific MSQ Plus Mass Detector, a single quadrupole mass spectrometer and Thermo Scientific Xcalibur software, the system detected marijuana cannabinoids, Δ9-tetrahydrocannabinol (THC), cannabinol and cannabidiol in the baked goods, a complex food matrix. The team delivered results with simpler sample preparation, no derivatisation and less instrument clean-up time.

The team extracted and separated cannabinoids on a Thermo Scientific Hypersil GOLD PFP column, and it showed that the cannabinoid standards eluted with good resolution at 4.1 minutes (cannabidiol), 5.1 minutes (THC) and 5.4 minutes (cannabinol). The extracts from a brownie and a cookie eluted at 5.1 minutes, testing positive for THC, demonstrating how little sample preparation is required for this method. It is simpler, faster and requires less sample material than traditional GC/MS techniques. Preparation time of ten minutes and a total run time of eight minutes shows the efficiency of UHPLC/MS detection.

The Accela™ is a high-speed chromatographic system that provides fast and efficient chromatographic separations over a range of flow rates and pressures. It features optimised system delay volumes for fast separations and flexible sample format. The innovative quaternary pump is capable of conventional and ultra-high pressures while the Thermo Scientific LightPipe technology provides maximum sensitivity and resolution.


Digital Edition

Lab Asia Dec 2025

December 2025

Chromatography Articles- Cutting-edge sample preparation tools help laboratories to stay ahead of the curveMass Spectrometry & Spectroscopy Articles- Unlocking the complexity of metabolomics: Pushi...

View all digital editions

Events

Smart Factory Expo 2026

Jan 21 2026 Tokyo, Japan

Nano Tech 2026

Jan 28 2026 Tokyo, Japan

Medical Fair India 2026

Jan 29 2026 New Delhi, India

SLAS 2026

Feb 07 2026 Boston, MA, USA

Asia Pharma Expo/Asia Lab Expo

Feb 12 2026 Dhaka, Bangladesh

View all events