The AllerGard NU-620 Dual Access Animal Transfer Station (ATS) employs a ‘push-pull’ pattern of airflow to provide clean air in the work zone for small animal cage caging. A supply impeller creates the ‘push’ in the transfer station by drawing air through a pre-filter and then propelling it past a supply HEPA filter and toward the work zone to create a clean air environment. From there, a powerful DC ECM motor underneath the work zone will ‘pull’ the air down through the air slots in the work surface. The air is then drawn past an exhaust pre-filter and pushed out an exhaust HEPA filter on the bottom of the unit. The Aeromax control system displays the performance of this unit by visualising the plenum pressure as a series of multi-coloured LED indicators. The NU-620 is designed as a dual-sided station to provide flexibility to vivariums with little room to manoeuvre.