Chromatography
Ion Chromatography: Fully Automated Inline Eluent Preparation
May 21 2009
Metrohm presents a system for fully automated inline eluent preparation. Based on Metrohm’s intelligent ion chromatography systems,
the new technique eliminates error-prone and labour-intensive eluent preparation steps while improving the accuracy of the chromatographic analysis. Using the 800 Dosino and the 849 Level Control as the only additional devices, Metrohm’s intelligent IC systems – the 850 Professional IC and the Compact IC family – can be easily extended to perform any inline eluent preparation. Controlled by the MagIC Net™ software, the 849 Level Control monitors the eluent level while the 800 Dosino performs all dosing and liquid handling tasks.
In a test, consecutive injections of a 250-μg/L standard over approximately 20 days revealed an excellent retention-time stability. After more than 800 consecutive injections, relative standard deviations for standard anions and cations were smaller than 0.55 and 0.41% respectively. In the case of a 24-hour sequence, retention-time precision for anions and cations was better than 0.09 and 0.08%, respectively. In short, Metrohm’s inline eluent preparation system increases the retention-time reproducibility and allows the continuous determination of anions and cations without manual eluent preparation.
Digital Edition
International Labmate Buyers' Guide 2024/25
June 2024
Buyers' Guide featuring: Product Listings & Manufacturers Directory Chromatography Articles - Enhancing HPLC Field Service with fast-response, non-invasive flowmeters - Digital transformatio...
View all digital editions
Events
Jul 03 2024 Gandhinagar, India
Jul 07 2024 Dublin, Ireland
Jul 20 2024 Denver, CO, USA
Jul 21 2024 Cape Town, South Africa
Jul 28 2024 San Diego, CA USA