Solving industry challenges for phosphorus containing compounds with Benchtop NMR
Oct 31 2024
Author: Elvis Wen, Robin J. Blagg on behalf of Oxford Instruments Magnetic Resonance
Free to read
Articles are free to download. Unlock the article to be shown more content, graphs and images.
In this application note, we present case studies highlighting how benchtop NMR characterises phosphorus containing compounds and materials across their lifecycle. Phosphorus naturally occurs in phosphate containing rocks, is active in living organisms and hence a cornerstone of life. Phosphorus containing compounds are important in industries including agriculture, chemical manufacturing, pharmaceuticals, and energy.
In agriculture, as a nutrient necessary for plant growth, phosphate fertilisers promote the development and maturation of plants. Phosphorus-containing pesticides and insecticides help improve the quality and yield of crops. In the chemical industry, phosphates and phosphides are contained in coatings, flame retardants, and detergents. In the field of medicine, phosphorus compounds are used as active ingredients (APIs) and excipients in the synthesis of antiviral and anticancer drugs and as biomolecule labels. In batteries, lithium-ion phosphate, and lithium hexafluoro-phosphate, are critical components in high power density current carrying electrolytes.
To manufacture high-quality phosphorus-containing products, it is necessary to characterise and analyse the target substances at all stages of production; to ensure that the product identity and purity is as required. The composition structure, purity and concentration of compounds containing phosphorous-31 is straightforward to analyse with benchtop NMR spectroscopy. With a natural abundance of 100%, a nuclear spin of one-half, and a high receptivity (relative to 13C) of 391; these properties make 31P one of the easiest nuclei to study by NMR. In addition, since the number of phosphorus atoms per molecule of phosphorus-containing compounds is generally small, the resulting spectra will be relatively simple. Cryogen free benchtop NMRs, can be widely applied to characterise phosphorus containing compounds across both R&D and manufacturing workflows.
This application note will present several examples of benchtop NMR 31P spectroscopy to assist in the characterisation of phosphorus-containing compounds.
Applications of Benchtop 31P NMR Spectroscopy
Detecting oxidation and degradation of raw materials
Phosphorus is most commonly present in compounds in +3 or +5 oxidation states. The different oxidation states may be difficult to distinguish with conventional analytical methods. For example, in Figure 1, spectra of four compounds are shown:
• _triphenylphosphine, PPh3;
• _triphenylphosphine oxide, OPPh3;
• _triphenylphosphite, P(OPh)3; and
• _triphenylphosphate, OP(OPh)3.
These compounds have similar molecular structures, with PPh3 & OPPh3, and P(OPh)3 & OP(OPh)3, only varying in their oxidation state and the presence of a single oxygen atom. In the common one-dimensional 1H or 13C{1H} NMR experiments spectra, these compounds are effectively indistinguishable. However, the compounds can clearly be distinguished in the 31P{1H} NMR spectrum, with the chemical shifts of the different compounds varying over a ca 150 ppm range.
If there are phosphorus-containing raw materials or products, that are prone to oxidation and deterioration, then 31P NMR spectroscopy is an effective method for detection and analysis of chemical degradation, potentially solving the problem with a single rapid measurement.
Analysis of Small Nucleoside Active Pharmaceutical Ingredients (APIs)
Compared with traditional small molecule and antibody drugs, small nucleic acid drugs intervene from the source of genes and have been described as ‘treating the symptoms and the root cause’. Hence they have become a promising new direction for drug development. Modified nucleosides and phosphoramidite monomers (Figure 2) are the core raw materials for the synthesis of small nucleic acid drugs, and their quality is one of the key links in the quality control strategy of small nucleic acid drugs.
A small nucleoside raw material manufacturer prepared four different phosphoramidite monomer products. One-dimensional 31P spectra of products N1, N2, N3 and N4 were acquired using the Oxford Instruments X-Pulse Broadband Benchtop NMR Spectrometer (Figure 3). It can be seen that the resonance signal of the target product, phosphoramidite, is located at a chemical shift of ca δP150 ppm. Although the four products are broadly similar in chemical composition, we can clearly observe the difference in the 31P splitting pattern in the spectra (Figure 2 inserts). Therefore, benchtop NMR can provide a direct and effective analytical method for the differentiation and identification of individual products.
In addition, the signal peaks of impurities or degradation products are clearly shown in the spectra of all four samples. The quantitative nature of NMR easily measure the relative content of the target product and impurities. Therefore, benchtop NMR enables evaluation of the quality, condition, and stability of the product. For example, there are more and stronger impurity signals in sample N1, indicating that the product has degraded more than the other three. Benchtop NMR can also be used as a quality control tool for the preparation and production of small nucleic acid drug products.
Battery electrolyte analysis
Low cost Lithium hexafluorophosphate, Li[PF6], is highly soluble and thermally stable in alkyl carbonate solvents, has strong ion-diffusion and migration capabilities. These advantages have led to its large-scale industrial production, and use as the most common lithium salt in commercial lithium-ion battery electrolytes. However, [PF6]− is sensitive to water, and trace moisture in the air is sufficient for it to decompose through hydrolysis. Therefore, the production and storage of Li[PF6] needs to be strictly controlled and tracked.
In the one-dimensional 31P NMR spectrum (Figure 4 above), the phosphorus atom of [PF6]− is coupled to six equivalent fluorine atoms and therefore splits to give a seven-peak signal, with peak intensities in a 1:6:15:20:15:6:1 ratio. After exposing [PF6]− to atmospheric moisture, a new signal of three peaks in a 1:2:1 ratio is observed in the one-dimensional 31P spectrum, in addition to the original seven peak signal (Figure 4).
Figure 3: One-dimensional 31P{1H} spectra of four phosphoramidite samples. The chemical shift ~150 ppm signal is enlarged in the inset plot.
This is due to the degradation of a portion of [PF6]− to form difluoro-phosphoric acid, OP(F)2(OH), with the splitting into three peaks caused by coupling to two equivalent fluorine atoms. Benchtop NMR is a powerful tool to identify electrolyte degradation products and the root causes of failure.
Additives both enhance the performance and extend the lifetime of Li-ion battery electrolytes. For example, nitrates are mainly used to improve cycling performance, sulphites to enhance low-temperature performance, and phosphate esters and fluorinated substances to improve flame retardancy and safety. Figure 5 shows the one-dimensional 31P spectra of phosphorus-containing additive samples from two different companies. Through the comparison of the spectra, it can be seen that the additives selected by the two companies have very different chemical compositions and structures.
NMR structural and compositional characterisation is key for R&D of new additives - for example tris(trimethyl-silane) phosphate (TMSP) and tris(trimethyl-silane) phosphite (TMSPi). Their main functions are to form a protective film, to reduce electrolyte decomposition, and remove water and acid. The one-dimensional 1H spectra of these two species are indistinguishable. However, there is a ca 300 ppm difference between the chemical shifts of TMSP and TMSPi in their 31P spectra (Figure 6).
Benchtop NMR spectroscopy is a powerful tool for the characterisation and analysis of battery electrolytes across their full lifecycle. The broadband X-Pulse system enables data from 31P NMR to be combined analysis of other chemical nuclei (inc. 7Li, 23Na, 11B, 19F, 13C, 1H) to give a comprehensive characterisation of all key electrolyte constituents.
Figure 6: One-dimensional 31P spectra of phosphorus additives TMSP (left) and TMSPi (right) in samples of battery electrolyte.
Improving chemical synthesis processes
A large chemical company provided two sets of four synthetic products to collect one-dimensional 31P{1H} spectra on the X-Pulse. The spectra of the first set of products, S1 and S2, are basically the same (Figure 7 left). The largest difference was of the signals in the range δP 160-170 ppm, where there were some more impurity peaks in product S2, in comparison to product S1.
The main peak of the other pair of samples, Y1 and Y2, is around δP 163 ppm (Figure 7 right).
However, the other signals of product Y1 are significantly stronger than those of Y2, especially the signal at ca δP 133 ppm.
The spectral data obtained from these experiments enable manufacturing process optimisation to reduce chemical impurities by changing synthesis conditions and processing methods. This example highlights how benchtop NMR phosphorus spectra enable interventions to increase product quality and yield and reduce cost of production.
Figure 7: One-dimensional 31P{1H} spectroscopy is used to evaluate chemical products and preparation processes.
Summary
Phosphorus-31 NMR spectroscopy is a direct and effective characterisation method for detecting phosphorus signals in compounds without interference from other elements. Across many industries and for academic research, it provides structural and chemical characterisation, quanitification of purity and concentration and mapping of synthesis reaction effectiveness.
The Oxford Instruments X-Pulse Broadband Benchtop NMR Spectrometer is easy to operate, and maintain. It’s capability to measure a wide range of nuclei in addition to 31P provides comprehensive characterisation of compounds and end products to improve the accelerate R&D and improve manufacturing efficiency.
Related Content
Accelerating pharmaceutical R&D with NMR - https://nmr.oxinst.com/pharma
NMR applications across the battery lifecylce - https://nmr.oxinst.com/application-detail/nmr-in-battery-analysis
Chemical reaction feedstock troubleshooting - https://nmr.oxinst.com/application-detail/chemical-reaction-feedstock-troubleshooting-with-benchtop-nmr
Quality control of raw chemical materials - https://nmr.oxinst.com/application-detail/quality-control-of-raw-materials-in-the-chemical-industry
X-Pulse benchtop NMR spectrometer - https://nmr.oxinst.com/x-pulse
Free to read
Articles are free to download. Please login to read this article or create an account.
Digital Edition
Lab Asia 31.6 Dec 2024
December 2024
Chromatography Articles - Sustainable chromatography: Embracing software for greener methods Mass Spectrometry & Spectroscopy Articles - Solving industry challenges for phosphorus containi...
View all digital editions
Events
Jan 22 2025 Tokyo, Japan
Jan 22 2025 Birmingham, UK
Jan 25 2025 San Diego, CA, USA
Jan 27 2025 Dubai, UAE
Jan 29 2025 Tokyo, Japan