Analysis of Total Sulfur according to ASTM D5453, with the Conditioned Syringe Injection option for the Vectra autosampler
Aug 28 2025
Author:
Sebastian Sanchez
on behalf of TE InstrumentsFree to read
Articles are free to download. Unlock the article to be shown more content, graphs and images.
Introduction:
At the beginning of 2024, TE Instruments (TEI) launched the Vectra autosampler, a new, automated solution for introducing liquid samples into the TEI combustion analyzers. This innovative autosampler incorporates advanced motion technology, offering an extended range of movements and setting the next stage for future developments in automated sample handling.
Initially, the Vectra autosampler supported two optional conditioned sample trays, allowing users to heat or cool sample vials based on their specific application needs. Now, with the newly developed syringe conditioning module, the Vectra sampler goes a step further by enabling precise heating and cooling of the syringe during liquid handling and sample analysis. This powerful feature significantly expands the system’s versatility, especially for applications involving challenging or volatile sample matrices.
Background:
In modern analytical laboratories, automation is essential for increasing throughput, ensuring reproducibility, and minimizing human error. Among the most integral components of automated systems is the autosampler, which is responsible for the precise and accurate delivery of liquid samples to analytical instruments. Despite its routine use, proper liquid handling within an autosampler system is often underestimated — yet it plays a decisive role in data quality, instrument performance, and overall operational efficiency.
Accurate liquid handling ensures that the correct sample volume is aspirated and dispensed without contamination, evaporation, or carryover. This is particularly vital in high-sensitivity analyses such as combustion analysis, where even trace discrepancies can lead to false results, compromised calibrations, or costly reruns. Furthermore, improper handling can lead to needle blockages, increased wear on mechanical components, and irregular sample injections, ultimately impacting instrument uptime and maintenance costs.
One of the major challenges in handling low-boiling, volatile samples is ensuring proper aspiration and injection. Today, hydrocarbon-based samples with extremely low boiling points (25 degrees Celsius) are increasingly used in Total Sulfur, Nitrogen, or Halogen analysis. These samples present a significant risk of evaporating from the syringe before injection into the combustion analyzer — a scenario that was previously a barrier to their use.
At the other end of the spectrum, many waxy samples or heavy oils are solid at room temperature but become liquid when warmed. This makes them suitable for analysis when a conditioned sample tray is used to preheat the vials. A conditioned syringe ensures that the sample remains in a liquid state during transfer from the vial to the injection point or during cleaning cycles. Keeping the syringe heated not only maintains sample integrity but also extends the syringe’s lifespan by reducing the risk of clogging.
This article explores the possibilities and benefits of using a conditioned syringe with the Vectra autosampler.
Vectra autosampler with conditioned syringe option:
The conditioned syringe module for the Vectra consists of two key components:
1. A conditioned parking station, used to thermally condition the syringe before and after injection.
2. A customized syringe injection head, featuring a specially designed heating and cooling buffer block.
During operation, the syringe—secured in a specialized holder—is placed in the conditioning block. Depending on the application requirements, it is either heated or cooled to a predefined setpoint. The system can accurately adjust the syringe temperature within a range of 10°C to 65°C.
Thanks to the precision-engineered buffer block around the syringe, the syringe reaches the temperature setpoint within 2 minutes and maintains the target temperature within ± 5°C for 10 minutes. This ensures optimal temperature stability throughout the injection, washing cycles, and fully conditions the syringe for subsequent use.
This setup enhances process reliability and repeatability, particularly in temperature-sensitive applications where the sample’s behavior must be tightly temperature-controlled.
Experimental:
To confirm the principle of a conditioned syringe injection, we prepared an Xplorer Total Sulfur/ Total Nitrogen analyzer (TN/TS) with a Vectra Autosampler, 2 conditioned trays, and the syringe conditioning option.
Prior to analysis of samples, the analyzer setup was calibrated with calibration samples (Number of injections, 5 times per concentration level). The standards used for calibrations were made out of Pyridine (N) and Thiophene (S) in Xylene and prepared by the Vectra liquids autosampler.
After running the calibration line, the system was calibrated with a coefficient (R2) of 0.999994 for Total Sulfur.
Given that the analyzer was configured with a dual detector setup, including a CLD Nitrogen detector (Measuring Nitrogen as Nitrogen Oxide), a comparable calibration curve for Total Nitrogen was incorporated into the experimental procedure. This yielded a calibration line with a coefficient of determination (R²) of 0.999798, indicating good linearity for Total Nitrogen.
A variety of different sample matrices were collected from renowned laboratories or commercially acquired. These samples were injected into the analyzer combination with and without the conditioned syringe option, to compare average concentrations and variations for RSD%.
The following Table shows the experimental matrix setup with variations:
Table 1: Overview of sample matrices and corresponding variations
For comparison, all samples were injected with and without the syringe conditioning option; in all cases, all sample vials were pre-conditioned in conditioned sample trays of the Vectra sampler prior to injection.
The room temperature while running this experiment was between 26 to 28 ℃, this was the temperature for the injected samples without the Syringe Conditioning option.
Each sample matrix was subjected to 10 repetitions of 50 L as the injection volume. The average concentration and Relative Standard Deviation (RSD) were calculated and validated.
Results:
The following table shows the results of the experiment (all results with A are injections with the conditioned syringe option, and B are injections using the standard syringe configuration).
Table 2: Overview of concentrations and variation results of the experiment
Graph 1: Comparison overlay of injections of CRM Winter-gasoline with conditioned syringe (top) and without conditioned syringe (bottom). Red plot lines represent Total Sulfur graphs, Blue plot lines represent Total Nitrogen graphs.
Graph 2: Comparison overlay of injections of Pentane with conditioned syringe (top) and without conditioned syringe (bottom). The Red plot lines represent Total Sulfur graphs, Blue plot lines represent Total Nitrogen graphs.
Conclusion:
The comparative analysis of a syringe conditioned injection or without, across various fuel and bio-based matrices, reveals that incorporating syringe conditioning significantly improves both measurement precision and, in many cases, accuracy.
Across nearly all samples, a conditioned syringe resulted in lower %RSD values for both Total Sulfur and Total Nitrogen, indicating improved repeatability. The difference was particularly pronounced in the Pentane matrix; an injection without cooling results in extremely high variability (%RSD >5% for Total Sulfur and Total Nitrogen), compared to injection with the conditioned option (%RSD of 1.82% and 1.75%, respectively). This suggests a strong dependency on syringe conditioning in volatile or low-concentration matrices.
Overall, the inclusion of syringe conditioning enhances analytical reliability, reduces uncertainty, and is recommended as a standard practice, particularly for sensitive or variable matrices.
Used system setup:
For the validation study, the combination of the Xplorer TN/TS with the Vectra autosampler was utilized. This setup holds multiple sample trays, each with 50 sample positions for 2 ml vials. Additionally, two optional sample trays with temperature conditioning (heating or cooling) can be included, and a syringe conditioning option, to enhance the configuration.
Table 3: System settings of the Xplorer TN/TS
Free to read
Articles are free to download. Please login to read this article or create an account.
Digital Edition
Lab Asia Dec 2025
December 2025
Chromatography Articles- Cutting-edge sample preparation tools help laboratories to stay ahead of the curveMass Spectrometry & Spectroscopy Articles- Unlocking the complexity of metabolomics: Pushi...
View all digital editions
Events
Jan 21 2026 Tokyo, Japan
Jan 28 2026 Tokyo, Japan
Jan 29 2026 New Delhi, India
Feb 07 2026 Boston, MA, USA
Asia Pharma Expo/Asia Lab Expo
Feb 12 2026 Dhaka, Bangladesh