• Digital Pressure Meter Especially Adapted for GC

Chromatography

Digital Pressure Meter Especially Adapted for GC

Specially adapted for GC, the v-kit Digital Pressure Meter from NLG Analytical enables easy and accurate measurement of gas pressure at inlets. A universal capillary inlet adaptor is provided along with an Agilent compatible nut and ferrule, for connection to capillary inlets on 5890,6850,6890 & 7890. Other GCs can be connected by using the manufacturer’s standard capillary inlet nut and ferrule. The pressure meter is supplied with a traceable certificate of calibration.

Used in combination with NLG’s v-kit software it is possible to also use the digital pressure meter to diagnose common GC pressure problems. Inlet pressure accuracy is a key parameter in the OQ of any GC system. The inlet flow-rate set in GCs with electronic pressure control is simply derived from inlet pressure in conjunction with carrier gas type (such as He, H2, N2), column dimensions and temperature, making accurate pressure control essential.

VKIT-DMA Digital Pressure Meter Specifications: accuracy: ±0.2% FS at 23°C/73°F; repeatability: ±0.1% FS@23°C/73°F; operating temp: 0 to 50°C; proof pressure: 2x FS; battery: 9 V (included); battery life: 125 hours using MN1604; Dimensions: 178 H x 83 W x 29mm D; weight: 330g includes batteries; and pressure connection: special adaptor for Agilent GC instruments included.


Digital Edition

LMUK 49.7 Nov 2024

November 2024

News - Research & Events News   - News & Views Articles - They’re burning the labs... Spotlight Features - Incubators, Freezers & Cooling Equipment - Pumps, Valves & Liquid Hand...

View all digital editions

Events

analytica China

Nov 18 2024 Shanghai, China

Pharma Asia

Nov 20 2024 Karachi, Pakistan

Turkchem

Nov 27 2024 Istanbul, Turkey

Smart Factory Expo 2025

Jan 22 2025 Tokyo, Japan

Instrumentation Live

Jan 22 2025 Birmingham, UK

View all events