Chromatography

Vacuum Ultraviolet Spectroscopy - Rob Morris

Author: Rob Morris

Free to read

This article has been unlocked and is ready to read.

Download

A variety of materials have spectral signatures in the vacuum ultraviolet (VUV) region of 10–200 nm. VUV spectroscopy is useful for biomedicine, semiconductor testing and other applications. Because the spectral response of standard silicon charge-coupled device (CCD) detectors drops off rapidly at wavelengths as long as 400 nm, and because oxygen and water absorb in the VUV spectral band, maintaining sufficient signal outside a vacuum is not possible. Nitrogen purging of the spectrometer helps to mitigate water and oxygen absorption in the VUV, but spectrometers with a vacuum purging feature require custom configuration and are too expensive and unwieldy for many common applications.

Testing the Maya2000 Pro Miniature Fibre Optic Spectrometer
The Maya2000 Pro Spectrometer (Figure 1) delivers high quantum efficiency and high dynamic range and is responsive in the deep UV (~185-300 nm). The Extra-Deep UV option extends spectral measurements to 153 nm. The spectrometer features a 101.6-mm focal length optical bench with a compact crossed Czerny-Turner design and a back-thinned CCD detector that offers superior UV and VUV performance. To minimise signal attenuation inside the optical path, the spectrometer bench is purged with nitrogen, achieving a robust signal for VUV experiments at wavelengths as low as 153 nm.

Experimental Conditions
To test performance, a Maya2000 Pro was configured with a high resolution, UVenhanced 2400g/mm holographic diffraction grating and a 5 μm slit. The backthinned detector (Hamamatsu S10420-1106) features a peak quantum efficiency of 75%, with UV quantum efficiencies as high as 50%. A custom magnesium fluoride (MgF2) glass window is placed over the detector for improved transmission over VUV wavelengths. The test sample for the experiment was a VUV deuterium lamp with a VUV-grade optical window, coupled directly to the spectrometer. We enclosed the source and spectrometer and purged the region with standard laboratory-grade nitrogen.

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

Lab Asia Dec 2025

December 2025

Chromatography Articles- Cutting-edge sample preparation tools help laboratories to stay ahead of the curveMass Spectrometry & Spectroscopy Articles- Unlocking the complexity of metabolomics: Pushi...

View all digital editions

Events

Smart Factory Expo 2026

Jan 21 2026 Tokyo, Japan

Nano Tech 2026

Jan 28 2026 Tokyo, Japan

Medical Fair India 2026

Jan 29 2026 New Delhi, India

SLAS 2026

Feb 07 2026 Boston, MA, USA

Asia Pharma Expo/Asia Lab Expo

Feb 12 2026 Dhaka, Bangladesh

View all events