• Method transfer from Waters Alliance HPLC systems to the latest Agilent UHPLC platform
    Figure 1: Method transfer from Waters Alliance HPLC 2695 to 1290 Infinity Binary and 1290 Infinity Binary with ISET. Sample: Pesticides; column: 4.6 x 100mm ZORBAX Eclipse Plus, 5µm; flow: 2mL/min, 25-95% ACN-Water.
  • Figure 2: The Agilent 1290 Infinity LC with ISET executes any legacy HPLC or latest UHPLC method while delivering the same chromatographic results – all through a single mouse click!
  • Figure 3: The wide power range of the Agilent 1290 Infinity LC ensures that both HPLC and UHPLC methods from narrow bore to standard bore columns can be executed.

Chromatography

Method transfer from Waters Alliance HPLC systems to the latest Agilent UHPLC platform

Dec 10 2012

By Christian Gotenfels
Agilent Product Manager for Liquid Chromatography


Instrument-to-instrument method transferability is an important topic for all laboratories throughout all industries, where HPLC and UHPLC methods are transferred between different departments and locations with different LC instruments. Especially in highly regulated environments such as quality control in the pharmaceutical industry, the transfer of LC methods is a challenge because any modification of the original method needs to be avoided.

Which parameters affect method transfer?
Design differences between LC instrumentation – such as power range, delay volume, mixing behaviour, temperature control, extra column volume and detector cell design – all affect the ability to transfer a method from one system to another. Especially the transfer of methods from older HPLC platforms such as Waters Alliance to the latest UHPLC technology platforms such as the Agilent 1290 Infinity LC could result in different retention times and chromatographic resolution.

The impact of delay volume and gradient mixing
The delay volume of an LC system determines how fast the gradient reaches the column. Further, the mixing behaviour influences the gradient profile. The Waters Alliance 2695 HPLC system is specified with a delay volume of 650 µL and the eluents are mixed together on the low pressure side of the pump. In contrast the delay volume of the high pressure mixing Agilent 1290 Infinity Binary LC system is <140µL. The consequences for method transfer are differences in retention times and in chromatographic resolution.

Agilent 1290 Infinity Binary LC with ISET for infinitely better instrument-to-instrument method transfer
Agilent’s ISET – Intelligent System Emulation Technology – makes the Agilent 1290 Infinity Binary LC the most adaptive UHPLC system available. It can execute any legacy HPLC or latest UHPLC method and deliver the same chromatographic results without the need to change the original method or modify the instrument hardware.

This new technology is based on two Agilent innovations:

  • Best in class performance of the Agilent 1290 Infinity Binary LC
    The Agilent 1290 Infinity Binary LC with its broad power range, unmatched flow and composition accuracy, ultralow delay volume and superior sensitivity, delivers the key requirements for the implementation of ISET.
  • The revolutionary emulation algorithm
    With exact knowledge about the system behaviour of the target LC instrument and the high accuracy of the Agilent 1290 Infinity Binary LC, ISET is able to create an emulation function, which delivers the same gradient conditions as the selected instrument. No shift in retention times, no change in chromatographic resolution.

Higher productivity and cost savings for both method development and QA/QC 
With 1290 Infinity Binary LC and ISET you can speed up your method development with UHPLC performance and then fine-tune your method by emulating the target system – and be confident that the method will run as intended. No more method transfer problems in QA/QC! Simply emulate the LC system on which the original method was developed – with a single mouse click. You can run your legacy methods with ISET and at the same time take full advantage of the UHPLC speed, resolution and sensi tivity of the Agilent 1290 Infinity Binary LC. No need to maintain your old legacy LC systems.

For more information about Agilent’s Intelligent System Emulation Technology (ISET), see www.agilent.com/chem/infinity-iset


Digital Edition

International Labmate Buyers' Guide 2024/25

June 2024

Buyers' Guide featuring: Product Listings & Manufacturers Directory Chromatography Articles - Enhancing HPLC Field Service with fast-response, non-invasive flowmeters - Digital transformatio...

View all digital editions

Events

EuCheMS Chemistry Congress

Jul 07 2024 Dublin, Ireland

HPLC 2024

Jul 20 2024 Denver, CO, USA

ICMGP 2024

Jul 21 2024 Cape Town, South Africa

ADLM 2024

Jul 28 2024 San Diego, CA USA

InaLab 2024

Jul 30 2024 Jakarta, Indonesia

View all events