Laboratory Products
How Pipette Tips Influence Results
Apr 08 2016
Often experimental data is not reproducible even though all variables within the experimental procedure withstood a review. But is that really so?
In many cases the assumption is made that the pipetting system, consisting of pipette and pipette tip, is not a source of error as long as the tip fits the pipette. However, even if this is the case, it does not speak to the precision and accuracy of the pipetting result. Therefore, with good reason, the international standard ISO 8655:2002 recommends the use of pipettes and tips from the same manufacturer.
Only under these conditions the client can be sure that each lot of manufactured pipette tips represents exactly the calibration result described in ISO 8655:2002. This is not the case for a non-system supplier; a non-system supplier does not have any system specifications to which he must refer. Hence, the user is forced to perform calibrations and, if necessary, adjustments of his pipette with the foreign tips, every time tips from a new lot number are used.
The results of a large scale study in which a broad array of factors was identified which may negatively influence pipetting results can be found in Application Note 354, ‘The Tip of the Iceberg – how tips influence results’ at www.eppendorf.com/eptips.
Digital Edition
International Labmate 49.6 - Sept 2024
September 2024
Chromatography Articles - HPLC gradient validation using non-invasive flowmeters Mass Spectrometry & Spectroscopy Articles - From R&D to QC, making NMR accessible for everyone: Putting NMR...
View all digital editions
Events
Sep 29 2024 Singapore
Oct 06 2024 Liverpool, UK
Oct 08 2024 Gothenburg, Sweden
Oct 09 2024 Birmingham, UK
Oct 09 2024 NEC, Birmingham, UK