CAV® 4.1
9725-D35
CAV® 4.1 Single-Bath Benchtop Kinematic Viscometer
The CAV 4.1 replaces CANNON's trusted, obsolete CAV-2100 viscometer.
In stock
9725-D35
CAV® 4.1 Single-Bath Benchtop Kinematic Viscometer
The CAV 4.1 replaces CANNON's trusted, obsolete CAV-2100 viscometer.
In stock
CANNON® CAV® 4.1 is a fully automated, single bath, benchtop viscometer with two 14-position sample handlers for unattended D445 processing. Each of the two tubes covers a 100-fold viscosity range at values between 0.5 mm2/s (cSt) and 10,000 mm2/s (cSt) from 40 °C to 150 °C (down to 20 °C with optional integral bath cooler).
The CAV 4.1 replaces CANNON's trusted, obsolete CAV-2100 viscometer.
Conforms to ASTM D445/446. Common applications include formulated oil analysis, hydraulic oil analysis, additive analysis, marine fuel testing, base stock analysis, light and heavy fuel testing, waxes/paraffin, crude oil testing, and glycols/polyols.
Viscometer Tube Configuration Information
Protect your investment and ensure your instrument continues to provide quality and reliable results for years to come with a CANNON Service Plan.
*Some upper viscosity measurements may be limited by test temperature and sample type. Ubbelohde-style tubes (for volatile products that are not a pure solvent and high melt waxes) are also available in 100-fold increments and fast run 10-fold increments from 0.5 mm²/s (cSt) to 10,000 mm²/s (cSt).
Unit Weight | 54 kg (130 lb) |
---|---|
Method | ASTM D445/D446, IP 71, ISO 3104/3105, SAE J300 |
Dimensions (W x D x H) | 36.0 cm × 66.0 cm × 72.0 cm (14.3 in × 26.0 in × 28.5 in) |
Shipping Dimensions (W x D x H) | 81.3 cm × 53.3 cm × 99.0 cm (32.0 in × 21.0 in × 39.0 in) |
Shipping Weight (With All Items) | 96.6 kg (213.0 lb) |
Maximum Throughput | 24 tests per hour |
Sample Capacity | 28 (2 x 14 positions), automated |
Viscosity Range | 0.5 mm²/s (cSt) to 10,000 mm²/s (cSt) in 100-fold increments (depending on viscometer tube selection). Fast run tubes are also available. |
Timing Resolution | 0.01 s (timing accuracy to ± 0.001 s) |
Temperature Range and Accuracy | 40 °C to 100 °C,± 0.01 °C (down to 20 °C with cooling option) |
Minimum Sample/Solvent Volume | 8 mL sample (as little as 3 mL with fast run tubes)/15 mL solvent per test |
Operating Conditions | 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II; Pollution Degree 2 |
Electrical Specifications | 100 Vac, 50/60 Hz; 115 Vac, 50/60 Hz; 230 Vac, 50/60 Hz; 1200 watts power consumption |
Compliance | CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc, 60 sec.); ROHS |