FI filters
(top)
< 0.15 pM (< 3 amol/well FITC, 384sv, 20 μL).
FI filters
(bottom)
< 1.0 pM (< 50 amol/well FITC, 384g, 50 μL).
FI monochromator
(top)
< 0.35 pM (< 7 amol/well FITC, 384sv, 20 μL).
FI monochromator
(bottom)
< 3.0 pM (< 150 amol/well FITC, 384g, 50 μL).
Dynamic Range: 8 decades in a single measurement.
FP
< 0.5 mP SD at 1 nM FITC (384sv, 20 μL).
HTRF®
(black and white microplates)
Reader Control Kit (Eu) after 18h (384sv, 20 μL).
Delta F > 880% (High Calibrator).
Delta F > 30% (Low Calibrator).
TRF
< 20 fM europium, 384, 80 μL (<1.6 amol/well).
LUM
< 0.4 pM (< 8 amol/well ATP, 384sv, 20 μL).
Dynamic range: 8 decades in a single measurement.
AlphaScreen® with laser.
< 5 pM (< 100 amol/well P-Tyr-100, 384sv, 20 μL).
ABS with spectrometer.
Selectable spectral resolution: 1, 2, 5, and 10 nm.
OD range: 0 to 4 OD; photometric resolution: 0.001 OD.
Accuracy: < 1% at 2 OD.
Precision: < 0.5% at 1 OD and < 0.8% at 2 OD.
Linearity: ≤ 0.8% at 2.0 OD.