Laboratory Products

Homogenisation and characterisation of cannabidiol (CBD) isolates

Apr 13 2021

Author: Barry Schubmehl on behalf of Fritsch GmbH

Free to read

This article has been unlocked and is ready to read.

Download

Solid isolates of CBD may be created as a non-homogenous ‘cake’ of material, or as a mixture of loosely associated granules comprising a wide particle size distribution. Creating material suitable for consumption or subsequent formulation in other products typically requires particle size reduction and/or homogenisation. Confirmation of particle size distribution of a given sample may be achieved via traditional sieving methods, or by using a particle size analyser.

Primary considerations when selecting a milling system for CBD homogenisation include:
• Batch size (throughput requirement kg/batch/day)
• Minimising losses (physical and/or chemical)
• Desired final particle size range
• Ease of cleaning to prevent batch cross-contamination

Homogenisation - small-scale

Small-scale homogenisation of CBD isolate may be achieved by low energy ball milling or by automated mortar and pestle. Larger batches may be processed in a continuous manner by using a Variable Speed Rotor Mill  with a Cyclone separator.

Figure 1: Inside of a Variable Speed Rotor Mill PULVERISETTE 14 classic line, showing fixed
collection vessel, rotor and sieve ring. The stainless steel Cyclone separator is shown on the right.

Configured as a continuous milling system, the PULVERISETTE 14 classic line with a Cyclone allows continuous feeding of material, and active removal of material from the rotor area - typically within a fraction of a second. Because the homogenisation occurs so quickly and the material is actively removed, there is no time for frictional temperature increase to occur that could cause chemical degradation. Product contact surfaces are stainless steel, or food-grade plastic vacuum hose for the Cyclone, and can all be cleaned thoroughly between batches to reduce the likelihood of cross-contamination. Variable speed motor and range of sieve rings, provides control of particle size output.
Used together with collection vessel, the PULVERISETTE 14 classic line and Cyclone separator allow continuous feeding and homogenisation of CBD material into the low micron range (consistency of talcum powder).

Homogenisation - large-scale

Production-scale homogenisation of CBD isolate into uniform powder can be achieved with a Universal Cutting Mill. Already considered the premier, precision milling solution for preparation of cannabis biomass for extractions, or pre-roll products, the cutting mill may be configured as a large-scale homogeniser to allow continuous processing of large batches of isolate material.
Suggested system settings:
• Between 2000 - 3000 rpm
• Rotor with notched edges and fixed knives
• Gap setting between cutting rotor & fixed knives: 1 - 2 mm
• Sieve cassette: 1, 2, or 4 mm (depending on nature of sample material and desired fineness)
• High-performance stainless steel Cyclone separator

Particle size analysis

Characterising the particle size distribution of a sample may be performed for the following reasons:

• To confirm that a milling system has achieved particles within an accepted range
• In Quality Control: As a criterion for release into the manufacturing process (raw material testing) or for product sale
• In R&D: To study the efficacy of a compound based on its particle size, and/or on the perceived quality of a formulated product

Particle size analysis by traditional sieving methods involves passing the material through a series of sieve pans with progressively smaller openings, and calculating the weight of material that is retained in each pan. Data obtained by using this method is limited by the number of different sieves that are used and can be confounded if and when particles stick to the metal surfaces.

Figure 3: The Laser Particle Sizer ANALYSETTE 22 NeXT.

Figure 4: The ANALYSETTE 22 was used to measure CBD isolate before & after homogenisation. Starting CBD sample (orange) contained a mixture of loose granules with a wide particle size distribution. Material homogenised using the Variable Speed Rotor Mill PULVERISETTE 14 classic line (purple) resulted in a particle size consistency similar to talcum powder.

Particle Sizing Instrumentation

Laser particle size analysers with reverse Fourier optical system, pioneered and patented by Fritsch over 35 years ago, have become a worldwide standard across a wide range of industries, for precise size characterisation of materials down into the low nanometre range. The Laser Particle Sizer ANALYSETTE 22 has been used successfully to characterise CBD isolates, using either a wet or dry dispersion method.

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

Lab Asia 31.2 April 2024

April 2024

In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...

View all digital editions

Events

Lab Indonesia

Apr 24 2024 Jakarta, Indonesia

Expomed Eurasia

Apr 25 2024 Istanbul, Turkey

AOCS Annual Meeting & Expo

Apr 28 2024 Montreal, Quebec, Canada

SETAC Europe

May 05 2024 Seville, Spain

InformEx Zone at CPhl North America

May 07 2024 Pennsylvania, PA, USA

View all events