AvCount Lite Particle Counter - SA1800-2

AvCount Lite Particle Counter, SA1800-2

The AvCount Lite portable automatic particle counter provides reliable results for determining the size and distribution of particles in light and middle distillate fuels, including aviation fuel and kerosene, biodiesel, low-viscosity oils, and hydraulic oils.

The particle counters simple operation makes testing bottle samples or online quick and easy, with results in under 3 minutes.

The AvCount Lite displays the cumulative, particles/ml and ISO 4406 cleanliness codes. Up to 600 measurements can be saved to the internal memory, downloaded to a PC/laptop, or printed via the internal printer.

Supplied with: sample bottles, sample probe, waste pipe, universal voltage power supply, USB cable, calibration certificate, and a USB memory stick containing handbooks and Partikel software for basic test method programming and data download.

Methods:  ASTM D975, ASTM D7619, IP 565, IP 630, CEN/TR 15367-1, CEN/TR 17548, DEF STAN 91-091, DEF STAN 91-86, IEC 60970, ISO 4406, JIS B 9932:2012, JIS B 9933:2021, JIS B 9934:2012

Features

  • ISO 11171 calibration
  • Cumulative Particles/ml
  • ISO 4406 Cleanliness Codes
  • Portable, compact instrument
  • Online/offline (bottle sample)
  • Stand-alone or PC controlled
  • Programmable via PC
  • Integrated printer
  • Battery power optional

.

Sample viscosity:
The AvCount Lite can analyze samples with kinematic viscosities of up to 64 mm2/s. When used in conjunction with the optional Sample Delivery System, the viscosity range is extended up to 200 mm/s.

Precision and accuracy:
AvCount Lite is fitted with a precision dual-piston pump. When measuring samples from a bottle, the pump draws an aliquot of the sample through the measurement cell at a consistent flow rate. When connected to a pressurized line, the pump provides accurate metering and control.

Protrend Software:
Optional Protrend Software SA1810-0 can be purchased to control an AvCount Lite from a personal computer or laptop. Local custom test methods can be created, tests run, and data downloaded. Protrend supports multiple AvCount Lite units, synchronizing the sampling and measurement to allow trend analysis.

Calibration and verification:
AvCount Lite utilizes 16-point calibration. Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated ‘Master’ AvCount Lite. Calibration can be carried out at a user site or at our authorized regional laboratories. Seta authorized training is required to perform a calibration. Verification can be performed using AvCount Verification Material.

PARTICLE SIZE RANGE ISO 11171: 4 µm(c) to >70 µm© (calibration for larger sizes available on request)
NUMBER OF MEASURING CHANNELS 6 size channels displayed on instrument, up to 16 programmable via PC 4 µm(c), 6 µm(c), 10 µm(c), 14 µm(c), 21 µm(c), 25 µm(c), 30 µm(c), 38 µm(c), 70 µm(c)
COUNTS PER MEASUREMENT (MAX) 60,000 per ml
SAMPLE VISCOSITY (MAX) 64 mm2/s (using an internal pump), 200 mm2/s (pressure fed @ 3 BarG) (SA1950-0 Sample Delivery System is available as an accessory)
SAMPLE TEMPERATURE RANGE Ambient 0 to 70 °C
RESULTS 900 result memory Date/time indexed (last 20 tests can be viewed on-screen) Download to PC/laptop port or ProTrend software and print via internal printer
SAMPLE FLOW RATE 30 ml/min ±5 ml/min
TOTAL SAMPLE VOLUME USED 80 ml for ASTM D7619 & IP 565, from 20 ml for other methods (includes flush cycles)
VOLTAGE 12 Vdc 3A with printer on, 1A with printer off Charger/mains adaptor 100 to 240 Vac, 50/60Hz input, 12 Vdc 2.5A output Optional 12 Vdc battery, and charger (SA1816-0)
SIZE (HXWXD) 25 cm x 14.5 cm x 35 cm
WEIGHT 7.5 kg
CCCN CODE Tariff 90278990