AN-RS-001

2025-02

Identifying polymers with Raman spectroscopy

Rapid, nondestructive determination of colored plastics


Summary

Despite its precision in analyzing polymers, Raman spectroscopy struggles to accurately identify colored plastics due to fluorescence interference from pigments and additives. Polymer identification results may be less reliable under these conditions.

Utilizing multiple excitation wavelengths, comprehensive reference libraries, and employing advanced techniques all improve the reliability of Raman analysis. Metrohm’s Raman solutions encompass all of these capabilities and provide fast, accurate polymer identification. This Application Note focuses on Metrohm’s advanced XTR® technology which improves identification of substances by eXTRacting the Raman signal from spectra with strong, occluding background fluorescence. Several types of colored polymers are examined, showing the benefits of using 785 nm Raman spectroscopy with the novel XTR fluorescence rejection technique.


Introduction

This Application Note describes the identification of poly(ethylene-vinyl acetate) – also known as PEVA, polystyrene (PS), and polypropylene (PP) in colored plastics with Raman spectroscopy.

Measurements with the handheld MIRA XTR Raman spectrometer require no sample preparation and provide immediate and unambiguous results. Rapid and nondestructive determination takes place through application of the XTR algorithm followed by an automated library search.


Experimental

All spectra were measured using MIRA XTR, at a wavelength of 785 nm and with Orbital Raster Scan (ORS™), in auto-acquisition mode. Occluding fluorescence initiates XTR algorithms automatically, returning a high-resolution, baselined spectrum of the material.

All samples in this study were common household and office materials, tested through direct contact using the intelligent Universal Attachment (iUA).

Metrohm’s Illicit and General Chemicals Library is a comprehensive spectral database of common materials for accurate identification of mixed samples. Each sample was identified using this library in MIRA Cal software. A range of polymer standards is displayed in Figure 1, illustrating their distinct fingerprints in the Raman spectrum.

Figure 1. Stacked Raman library spectra for a wide range of polymer samples demonstrate how distinct the Raman fingerprint peaks are for each material.

Results

The Raman spectra of various items composed of unknown polymers in different colors were collected (Figure 2) and checked against the MIRA Illicit and General Chemicals Library.

Figure 2. Resolved, stacked Raman spectra of various strongly colored plastics.

Results display on the device within seconds and include the name, CAS number, and HQI (Hit Quality Index) – a measure of the strength of correlation between sample and library spectra. Figure 3 demonstrates the accuracy of data collected from even strongly colored samples by comparing sample and library spectra.

Figure 3. A direct comparison of experimental and library spectra demonstrates the accuracy and resolution of both spectra and results. Each plastic sample matched to a single library spectrum with high HQI values.

Conclusions

Metrohm has unique solutions that overcome the problem of fluorescence while retaining all the benefits of 785 nm interrogation — fast, accurate, nondestructive identification of colored polymers.